Operational specification
Measuring range:
-40 – 85ยฐC and Open/Closed
Resolution:
0.1ยฐC
Accuracy:
0.5ยฐC (-5 – 75 ยฐC)
Measuring Frequency*:
Every 3 sec
Report Frequency*:
Reports every 2 min. Or immediately if trigger for critical data transmission is reached, see below
Trigger for Critical Data Transmission*:
2ยฐC or change in status open/closed loop
Operating Environment:
Ambient temperature: -40 – 85 ยฐC
Relative humidity: 0-100%
Altitude: < 2000m above sea level
Pollution degree: 4
IP Grade:
IP 67, wet conditions, indoor use
Radio Frequency:
863-870 MHz / 902-928 MHz
Battery Type:
Li-SOCl2, 3.6V
Expected Operating Time**:
Up to 10 years
* Adjustable on request
** Depends on measurement frequency, amount of critical data transmissions and ambient temperature
Physical specification
Materials:
POLYblend 65 FS / TPU
Dimensions LxWxH:
37x23x14mm
Regulatory
Certifications:
CE marking
UK CA
FCC
Industry Canada
Directives/Standard:
IEC 61010-1:2010
FCC Part 15C
RED 2014/53/EU Radio